General
| Measuring principle |
Ultrasonic
time difference
correlation principle and doppler |
Flow
velocity range
|
0.01...25m/s
0.025 cm/s |
Resolution
|
0.15
% of measured value |
| Repeatibility |
± 0.015
m/s Volume flow |
Accuracy
|
± 1
... 3 % of measured value
depending on application,
±0.5 % of measured value withprocess calibration
Flow velocity ± 0.5 % of measured value |
Turn
down ratio
|
1/200 |
Gaseous
and solid
content of medium |
< 10
% of volume |
Flow
transmitter
Enclosure
|
Portable |
| Degree
of protection |
IP
54 according EN 60529 |
| Operating
temperature |
-10...60 ºC(14...140 ºF) |
| Housing
material |
Aluminium,
powder coated |
| Flow
channels |
2 |
| Power
supply |
Internal
rechargeable battery
6 V/4 Ah or external power
supply 9...15V DC |
| Operating
time |
>14
h with fully charged battery |
| Display |
2
x 16 digit LCD, dot matrix,
backlit |
| Dimensions |
H
118 x W 276 x D 310mm
(with handle) |
| Power
consumption |
< 2.5
W in measurement mode |
| Signal
damping |
0...60
s, configurable |
| Response
time |
1
s, 70 ms optional |
| Measuring
cycle |
100
... 1000 Hz, single channel |
| Calculation
functions |
Average/difference/sum |
Quantity
and units of measurement
| Volumetric
flow rate |
m³/h,
m³/min, m³/s, l/in, l/min, l/s,
USgph, bls/d (barrels per day) |
| Flow
velocity |
m/s,
inch/s |
| Mass
flow rate |
g/s,
t/h, kg/in, kg/min |
| Volume |
m³,
l, gal (gallons) |
| Mass |
g,kg,t |
| Heat
flow |
W.
kW, MW (only with heat
quantity measurement option) |
| Heat
quantity |
J.
kJ, MJ (only with heat quantity measurement option) |
Internal
data logger
| Storage
capacity |
approx.
27,000 (optional
> 100,000) measuring values |
| Logging
data |
All
measured and totalised
values, parameter sets |
Communication
| Serial
interface |
RS
232 |
Data
|
Instantaneous
measured value,
parameter set and configuration,
logged data |
Software
Data
| Functionality |
Downloading
of measured
values/parameter set, graphical
presentation, list format, export
to third party software, on-line
transfer of measured data |
| Operating
systems |
Windows 3.11,
95, 98, NT |
|
| Process
inputs
|
Galvanically
isolated from main
electronics |
| Temperature |
PT
100, four-wire circuit,
measuring range - 50...400 ºC |
| Current |
0...20mA;
Ri=50 ohm |
| Voltage |
0...1
V; Ri = 1 Megaohm |
| Process
outputs |
Galvanically
isolated from main electronics |
| Current |
0/4...
20 mA; passive (Uext< 24 V)
or active (Rext < 500 ohm) |
| Voltage |
0...1
V or 0...10V, Ri=500 ohm |
| Frequency |
0...1
kHz or 0 10 kHz; (OC) |
| Digital
(pulse, status) |
Totaliser
value 0.01 1000 / unit;
width 80...1000 ms; (OC/ Reed)
Reed = Reed-NO contact
(300 V /0.5 A)
OC = Open-Collector |
Clamp-on
sensors
Type M2N, M2E
Rated (possible)
diameter range
| Dimensions |
(50)
100...6500 mm
60 x 30 x 34 mm |
| Material |
Stainless
steel |
| Temperature
range |
Type
M2N:
-30 ... 130 ºC (-22 ... 266 ºF)
TypeM2E:
-30 ... 200ºC (-22 ... 392 ºF), for
short periods up to 300 ºC (572 ºF) |
| Degree
of protection |
IP
65 acc. EN 60529, IP 68 optional |
Type
Q3N, Q3E
Rated (possible)
diameter range
| Dimensions |
(10)
25...400 (1000) mm |
| Material |
43x18x22mm |
| Temperature
range |
Stainless
steel
Type Q3N:
-30...130 ºC (-22...266 ºF)
Type Q3E:
-30...200 ºC (-22...392 ºF),
for short periods up to |
| Degree
of protection |
300 ºC
(572 ºF)
IP 65 acc. EN 60529, IP 68 optional |
Special
Clamp-on sensors
| Type
S2N |
For
very small pipe
diameters 6 ... 40 (100) mm |
| Other
types |
On
request |
Wall
thickness measurement
| Measuring
range |
1.0...200
mm |
Resolution
|
0.01
mm |
| Linearity |
0.1
mm |
| Temperature
range |
Standard
version -20 ... 60 °C
High temperature version
0 ... 200 °C, for short periods up to 540 °C |
Accessories
-External
power supply 230 V, 50Hz/12 V, 1.2 A; IP 30
-Car power adapter 12 V, 2 A
-Leather carving case 330 x 340 x 220 mm
-Cable extension 3 m, 5 m, 10 m or 20 m
-Sensor positioning rail for sensors type Q3, stainless steel V2A
-External printer, ink jet 192 dpi |
|