General
Measuring principle Ultrasonic time difference corre-lation principle and doppler
Flow velocity range
0.01...25m/s
Resolution
0.025 cm/s
Repeatibility 0.15 % of measured value ±
0.015 m/s
Accuracy
Volume flow
± 1 ... 3 % of measured value
depending on application,
± 0.5 % of measured value with process calibration
Flow velocity ± 0.5 % of measured value
Turn down ratio
1/200
Gaseous and solid
content of medium
< 10 % of volume

Flow transmitter
Enclosure
Field transmitter housing
Degree of protection IP 67 according EN 60529
Operating temperature -10...55ºC(14...140 ºF)
Housing material Aluminium, powder coated
Flow channels 2
Power supply 100...240V AC/ 9...18V DC/
18...36V DC/ 36...72V DC
Display 2 x 16 digit LCD, dot matrix,
backlit, 2 X status LED
Dimensions H 213 x W 129 x D 170mm
(without cable glands and
supports)
Keyboard 5 keys, external access via
magnet
Power consumption < 15 W
Signal damping 0...60 s
Response time 1 s, 70 ms optional
Measuring cycle 100 ... 1000 Hz, single channel
Calculation functions Average/difference/sum
Protection concept
Flameproof (d), intrinsic safety
(i), increased safety (e)
Certification code Flameproof (d), intrinsic safety
(i), increased safety (e)

Quantity and units of measurement
Volumetric flow rate m³/h, m³/min, m³/s, l/in, l/min, l/s,
USgph, bls/d (barrels per day)
Flow velocity m/s, inch/s
Mass flow rate g/s, t/h, kg/in, kg/min
Volume m³, l, gal (gallons)
Mass g,kg,t

Communication
Serial interface RS485

Software Data 
Functionality Downloading of measured
values/parameter set, graphical
presentation, list format, export
to third party software, on-line
transfer of measured data
Operating systems Windows™ 3.11, 95, 98, NT
Current 0/4... 20 mA; passive (Uext< 24 V)
or active (Rext < 500 ohm)
Current 0/4... 20 mA; passive (Uext< 24 V)
or active (Rext < 500 ohm)
Frequency 0...1 kHz or 0 10 kHz; (OC)
Digital (pulse, status) 0...1 kHz or 0 10 kHz; (OC)
2 X totaliser value 0.01 1000 / unit;
width 80...1000 ms; (OC/ Reed)
Reed = Reed-NO contact
(300 V /0.5 A)
OC = Open-Collector

Clamp-on sensors
Type Q4N-Ex, M4N-EX (for use in hazardous areas
zone 1 and 2) Rated (possible)
diameter range

Type Q4N-Ex
(10) 25...400 (1000) mm
Type M4N-Ex
(50)100...3000 mm
Dimensions
(50) 100...6500 mm
60 x 30 x 34 mm
Material Stainless steel
Temperature range -20...120 ºC
Degree of protection IP 65 acc. EN 60529, IP 68 optional
Protection concept Encapsulation
Certification code EEx m II T4-T6
IBExU 98 ATEX 1012X